Please use this identifier to cite or link to this item: doi:10.22028/D291-48423
Title: Keratometry and the Central Radius with Aspheric Corneal Surfaces
Author(s): Langenbucher, Achim
Wendelstein, Jascha
Cayless, Alan
Hoffmann, Peter
Szentmáry, Nóra
Language: English
Title: Current Eye Research
Volume: 51 (2026)
Issue: 4
Pages: 397-403
Publisher/Platform: Taylor & Francis
Year of Publication: 2025
Free key words: Keratometry
central corneal radius
conoid surface model
corneal asphericity
DDC notations: 610 Medicine and health
Publikation type: Journal Article
Abstract: Purpose: To develop a simple concept for deriving the central corneal radius (RC) from the keratometry (K) or simulated keratometry (SimK) radius (RS) and corneal asphericity (QC), and to show the differences RC-RS based on a large dataset of measurements from a modern anterior segment tomographer. Methods: Comparing the local slopes of surface representations with a conoid and a reference sphere at the location of a keratometry measurement (diameter KD), RC and RC-RS can be derived as a function of RS, QC and KD. The differences RC-RS were evaluated for a large dataset containing measurements from the Casia 2 tomographer made before cataract surgery. Results: Depending on QC and KD, RC could deviate from the measured RS by up to 0.1 mm. With prolate corneas RS overestimates RC, and with oblate corneas RS underestimates RC. For example, with a typical cornea with RS = 7.7 mm and QC = -0.22 and a keratometer measuring at KD = 3 mm the central cornea is 0.0322 mm steeper (RC = 7.6678 mm) compared to keratometric measurement. Based on the dataset, the 95% confidence interval of RC-RS with KD = 3 mm was −0.0976 to +0.0265 mm. Conclusions: For corneal representations with a conoid surface or where central corneal radius is required e.g. for paraxial calculations, the keratometric radius of curvature could easily be converted to the central radius based on corneal asphericity and the keratometric zone diameter. With large positive or negative values of QC the differences RC-RS could be clinically relevant.
DOI of the first publication: 10.1080/02713683.2025.2596954
URL of the first publication: https://doi.org/10.1080/02713683.2025.2596954
Link to this record: urn:nbn:de:bsz:291--ds-484235
hdl:20.500.11880/42345
http://dx.doi.org/10.22028/D291-48423
ISSN: 1460-2202
0271-3683
Date of registration: 4-Aug-2026
Faculty: M - Medizinische Fakultät
Department: M - Augenheilkunde
Professorship: M - Univ.-Prof. Dr. Dipl.-Ing. Achim Langenbucher
M - Prof. Dr. med. Nóra Szentmáry
Collections:SciDok - Der Wissenschaftsserver der Universität des Saarlandes

Files for this record:
File Description SizeFormat 
Keratometry and the Central Radius with Aspheric Corneal Surfaces.pdf3,28 MBAdobe PDFView/Open


This item is licensed under a Creative Commons License Creative Commons